Product Overview
The TPS20 Series Non-indicating Pressure Transmitters offer excellent corrosion resistance with stainless steel housing and high accuracy of 0.3% F.S. They are designed for measuring gauge, absolute, and compound pressures in liquid, gas, and oil applications, excluding corrosive environments for stainless steel type 316. Available in head type, DIN connector type, and connector cable type, these transmitters feature user-friendly functions like built-in zero-point and span adjustment (head type). They are suitable for industrial environments with a temperature range of -10 to 70 and humidity up to 95% RH.
Product Attributes
- Brand: G-67 TPS20 Series
- Material: Stainless steel housing, Sealing/diaphragm/connection: stainless steel type 316, O-ring: fluoro rubber
- Certifications: Approval
- Protection Circuit: Reverse polarity protection circuit
- Case Structure: Drip-proof structure
Technical Specifications
| Specification | Details |
|---|---|
| Product Type | Non-indicating Pressure Transmitter |
| Model Series | TPS20 |
| Pressure Types | Gauge pressure, Absolute pressure, Compound pressure |
| Rated Pressure Range | 0 to 0.2 to 350kgf/cm (Gauge) 0 to 1.0 to 35kgf/cm (Absolute) -760mmHg to 0 to 35kgf/cm (Compound) |
| Max. Pressure Range | 300% of max. span |
| Measured Materials | Liquid, gas, oil (except corrosive environment of stainless steel type 316) |
| Power Supply | 15-35VDC |
| Permissible Voltage Range | 90 to 110% of rated voltage |
| Current Consumption | Max. 50mA |
| Response Time | Max. 100ms |
| Current Output | DC4-20mA |
| Linearity | 0.3% F.S. (-10 to 50), 0.5% F.S. (50 to 70) |
| Hysteresis | 0.3% F.S. |
| Temp. Zero Shift | 0.03% F.S. |
| Temp. Span Shift | 0.03% F.S. (at 25) |
| Load Resistance | Max. 600 |
| Insulation Resistance | Over 100M (at 500VDC megger) |
| Dielectric Strength | 500VAC 50/60Hz for 1 minute |
| Vibration | 1.5mm amplitude at frequency of 10 to 55Hz (for 1 min) in each X, Y, Z direction for 2 hours |
| Shock | 95m/s |
| Tightening Torque | Industrial plug over 5N |
| Pressure Port | G3/8" (standard), R3/8", R1/2" |
| Ambient Temperature | -10 to 70 (operation/storage) |
| Ambient Humidity | 5 to 95% RH (operation/storage) |
| Weight (approx.) | 350g (approx. 320g for unit only, based on head type) |
| Connection Types | Head type, DIN connector type, Connector cable type |
| Connection Cable Models | CID3-2, CID3-5, CLD3-2, CLD3-5 (sold separately) |
| Environmental Conditions | Indoor/Outdoor (within rated conditions), Altitude max. 2,000 m, Pollution Degree 2, Installation Category II |
Dimensions
Head type: M12, 14.8, 40 mm (approx.)
DIN connector type: M12, 14.8, 32 mm (approx.)
Connector cable type: M12 Connector, 14.8, 27.5 mm (approx.)
CLD3-2 / CLD3-5 & CID3-2 / CID3-5 Cables: L(m) - 2 or 5 meters
Connection Diagrams
Main Circuit (2-wire):
- Power: 24VDC (brown wire)
- Output: DC4-20mA (blue wire)
- Connections: +V (Pin 1), GND (Pin 2) for DIN connector type; + (Terminal 24VDC), - (Terminal Output)
Head type Terminal Connections:
- Pin +: +
- Pin 2: N.C.
- Pin F.G.: F.G.
- Pin -: -
DIN connector type Pin Functions:
- Pin 1: +
- Pin 2: N.C.
- Pin 3: N.C.
- Pin 4: F.G.
2411220112_Autonics-TPS20-G28P8-00_C7073353.pdf
Please Use Our Online Inquiry Contact Form Below If You Have Any Questions, Our Team Will Get Back To You As Soon As Possible