AO-07 Oxygen Sensor
The AO-07 Oxygen Sensor is an electrochemical sensor specifically designed for measuring oxygen concentration in medical equipment. Featuring a molded body design, it offers fast response times and a long service life. This sensor requires no external power for operation and comes factory-calibrated with professional accuracy and temperature compensation. It is suitable for use as a sensing component in oxygen analyzers, including oxygen concentrator control devices, medical ventilators, anesthesia equipment, and incubators.
Key Features:
- Full-range linear output
- No external power required
- Temperature compensated
- Fast response
- Accurate and reliable
- Strong anti-interference capability
Product Attributes
- Brand: ASAIR (implied by copyright)
- Model: AO-07
- Material: White ABS (housing)
- Interface: RJ11-4P4C
Technical Specifications
| Parameter | Specification |
|---|---|
| Measurement Principle | Partial Pressure Electrochemical |
| Output Voltage (in air) | 9~13 mV |
| Measurement Range | 0~100 % |
| Response Time (T90) | <15 s |
| Accuracy | 1 % |
| Linearity | Full-range linear |
| Temperature Compensation (O2 equivalent) | <2% (0~40 ) |
| Load Resistance | 10 k |
| Housing Material | White ABS |
| Weight | Approx. 40 grams |
| Operating Temperature Range | 0~50 |
| Operating Pressure Range | 0.5~2.0 Bar |
| Operating Humidity Range | 0~99% RH (non-condensing) |
| Long-term Drift in 100% O2 (voltage decay/year) | <5% |
| Service Life (1.5106 % O2 hours @ 20 ) | 1.5106 % O2 hours |
| Service Life (0.8106 % O2 hours @ 40 ) | 0.8106 % O2 hours |
| Shelf Life | 13 months from date of shipment |
Dimensions
Refer to Figure 2 for AO-07 external dimensions (Unit: mm, Tolerance: 0.15 mm).
Installation and Usage
Installation: Hand-tighten the sensor to ensure good airtightness. Do not use wrenches or similar mechanical tools to avoid damaging the sensor threads.
Storage and Use: Avoid exposure to high concentrations of organic solvent vapor during storage, installation, and operation. If using a sensor with a printed circuit board (PCB), clean the PCB with a degreaser before installation to prevent flux residues from volatilizing and blocking the sensor's breathable membrane. Do not use organic solvents on the sensor housing, as they may cause plastic cracking.
Cleaning: If the sensor housing is contaminated, clean with distilled water and allow to air dry. Do not use steam sterilization or expose the sensor to environments containing chemicals like ethylene oxide or hydrogen peroxide for extended periods.
Recommended Circuit
Refer to Figure 3 for the AO-07 recommended application circuit diagram. The formula for calculating oxygen concentration in the measured environment is: Oxygen Concentration = (Ax - A0) * 20.9 / ((A1 - A0) * 100%).
Pin Definition
Refer to Figure 4 for the AO-07 pin definition diagram.
Quality Assurance
Guangzhou Aosong Electronics Co., Ltd. provides a quality guarantee for its products for 13 months from the date of shipment. This guarantee is based on the technical specifications in the product datasheet published by Aosong Electronics. If a product is found to be defective within the warranty period, the company will provide free repair or replacement.
Disclaimer: This product is not intended for use in safety protection devices, emergency stop equipment, or any other application where product failure could result in personal injury, unless specifically authorized. Refer to the product manual before installation, handling, use, or maintenance. Failure to comply with recommendations may result in serious personal injury or death.
2205111530_Aosong-AO-07_C3012640.pdf