Online Service

Online Service

Contact Person
+86 136 6733 2386
quality Hydrogen Sulfide Sensor Amphenol EC4-1000-H2S Featuring Linear Response and Operating Humidity Range factory
<
quality Hydrogen Sulfide Sensor Amphenol EC4-1000-H2S Featuring Linear Response and Operating Humidity Range factory
quality Hydrogen Sulfide Sensor Amphenol EC4-1000-H2S Featuring Linear Response and Operating Humidity Range factory
>
Specifications
Measurement Range:
0ppm~1000ppm
Voltage - Supply:
-
Operating Temperature:
-20℃~+50℃
Interface:
-
Mfr. Part #:
EC4-1000-H2S
Key Attributes
Model Number: EC4-1000-H2S
Product Description

Product Overview

The EC4-1000-H2S is a Hydrogen Sulfide (H2S) electrochemical sensor designed for performance and reliability. It offers a measurement range of 0 - 1000 ppm with a sensitivity of 50 - 130 nA/ppm H2S. The sensor exhibits a linear response across its range and a fast t90 response time of less than 45 seconds. It is suitable for operation within a temperature range of -20 to +50 C and humidity levels of 15 - 90% RH (non-condensing). This RoHS compliant device is intended for use in various gas sensing applications.

Product Attributes

  • Brand: SGX Sensortech
  • Model: EC4-1000-H2S
  • Label Colour: Silver
  • RoHS Compliant: Yes
  • Origin: Registered in England

Technical Specifications

Specification Value
PERFORMANCE Range 0 - 1000 ppm
Sensitivity 50 - 130 nA/ppm H2S
Zero in air at 20 C <8 ppm H2S
Resolution 1 ppm H2S
Expected lifetime 2 years
Linearity linear across range
Response time, t90 <45 s
Long-term output drift change in output per month
OPERATING CONDITIONS Temperature range -20 to +50 C
Temperature coefficient <0.5% signal per C
Operating humidity 15 - 90% RH (non-condensing)
Pressure range 90 - 110 kPa (nominal)
Operating circuit see Electrochemical Toxic Sensor Application Note
Applied potential 0 V
Storage life 6 months in original packaging (0 - 25 C)
Weight 6 g
Dimensions see outline

Cross-Sensitivity Data

GAS Concentration (ppm) Sensor Response (As ppm H2S)
Carbon Monoxide 300 0
Sulfur Dioxide 5 0
Nitric Oxide 35 <3
Nitrogen Dioxide 5 0
Hydrogen 1000 0
Isobutylene 100 0

Important Note

Do not solder to the connector pins as this may damage the sensor and thereby invalidate the warranty. Details on recommended connector pins can be found in the Frequently Asked Questions within the Gas Sensor section of the SGX website.