Specifications
Measurement Range:
0ppm~1000ppm
Voltage - Supply:
-
Operating Temperature:
-20℃~+50℃
Interface:
-
Mfr. Part #:
EC4-1000-H2
Key Attributes
Model Number: EC4-1000-H2
Product Description

Product Overview

The EC4-1000-H2 is a hydrogen electrochemical sensor designed for measuring hydrogen gas concentrations. It offers a performance range of 0-1000 ppm with a maximum overload capacity of 2000 ppm. The sensor provides a sensitivity of 10-30 nA/ppm H2 and a zero in air at 20 C of less than 20 ppm H2. It features a resolution of 10 ppm H2 and operates linearly across its specified range. The response time (t90) is less than 70 seconds, and it exhibits a long-term output drift of less than 2% signal per month. This device is RoHS compliant.

Product Attributes

  • Brand: SGX Sensortech
  • Model: EC4-1000-H2
  • Label Colour: Silver
  • Warranty: 2 years
  • RoHS Compliant: Yes
  • Registered in England: No.08067077

Technical Specifications

Specification Value
Performance Range 0 - 1000 ppm
Maximum Overload 2000 ppm
Sensitivity 10 - 30 nA/ppm H2
Zero in air at 20 C <20 ppm H2
Resolution 10 ppm H2
Linearity Linear across range
Response time, t90 <70 s
Long-term output drift <2% signal/month
Temperature range -20 to +50 C
Operating humidity 15 - 90% RH (non-condensing)
Pressure range 90 - 110 kPa (nominal)
Applied potential 0 V
Storage life 6 months in original packaging (0 - 25 C)
Weight 5 g approx.
Dimensions See outline

Cross-Sensitivity Data

Gas Concentration (ppm) Sensor Response (As ppm H2)
Carbon Monoxide 50 200
Hydrogen Sulfide 24 0
Sulfur Dioxide 5 0
Nitric Oxide 35 <10
Nitrogen Dioxide 5 0
Ethene 100 80
Chlorine 10 0