Specifications
Operating Temperature:
-20℃~+50℃
Mfr. Part #:
EC4-20-PH3
Key Attributes
Model Number: EC4-20-PH3
Product Description

Product Overview

The EC4-20-PH3 is a Phosphine Electrochemical Sensor designed for accurate detection of phosphine gas. It offers a performance range of 0 to 20 ppm with a maximum overload capacity of 100 ppm. The sensor exhibits a sensitivity of 800-2000 nA/ppm PH3 and a low zero output in air at 20 C (<0.5 ppm PH3). It provides a resolution of 0.1 ppm PH3 and operates linearly across its specified range. With a response time (t90) of less than 60 seconds and long-term output drift of less than 2% signal/month, this sensor is suitable for applications requiring reliable and stable phosphine monitoring. It is designed to be RoHS compliant and comes with a 2-year warranty.

Product Attributes

  • Brand: SGX Sensortech
  • Model: EC4-20-PH3
  • Label Colour: Silver
  • RoHS Compliant: Yes
  • Warranty: 2 years
  • Registered in England: No.08067077

Technical Specifications

Specification Value
Performance Range 0 - 20 ppm PH3
Maximum Overload 100 ppm PH3
Sensitivity 800 - 2000 nA/ppm PH3
Zero in air at 20 C <0.5 ppm PH3
Resolution 0.1 ppm PH3
Linearity Linear across range
Response time, t90 <60 s
Long-term output drift <2% signal/month
Temperature range -20 to +50 C
Operating humidity 15 - 90% RH (non-condensing)
Pressure range 90 - 110 kPa (nominal)
Applied potential 0 V
Storage life 6 months in original packaging (0 - 25 C)
Weight 5 g approx.
Cross-Sensitivity: Hydrogen Sulfide (15 ppm) 0 ppm PH3
Cross-Sensitivity: Sulfur Dioxide (5 ppm) <0.9 ppm PH3
Cross-Sensitivity: Carbon Monoxide (1000 ppm) 0 ppm PH3
Cross-Sensitivity: Ammonia (50 ppm) 0 ppm PH3
Cross-Sensitivity: Hydrogen (1000 ppm) 0 ppm PH3
Cross-Sensitivity: Ethene (100 ppm) 0 ppm PH3